Ultra-Low-Power...
public project

Description

This analog front-end design is composed of four blocks. The Low-Power Low-Noise Amplifier (LNA) [1] simultaneously filters the input DC signal and amplifies it. Usually, electrodes output an unknown DC offset, canceled by the LNA; otherwise, it would be amplified together with the input signal and saturate the output.

After the first stage gain, we need a variable gain stage since the biomedical signals have a varying amplitude due to the electrode placement and the user's characteristics. A Variable Gain Amplifier (VGA) will implement the variable gain stage, based on the non-inverting operational amplifier with variable resistors in its feedback loop. This operational amplifier should have a large output voltage excursion and be capable of driving a relatively large load as efficiently as possible, so it should have a class AB output stage [2]

Finally, the VGA output signal must be filtered to limit the noise and interference from unwanted frequencies. Since the biomedical signals usually have a very low upper cutoff frequency, the filters should be made of very low transconductance amplifiers [3] to implement very large time constants with reasonably sized integrated capacitors. Additionally, the filter output must be as linear as possible to minimize distortion, so its transconductors must have improved linearity [4].

All these analog circuit blocks have in common the need for a biasing current. We chose a resistorless self-biased current reference source (SBCS) [5] since integrated resistors exhibit a more considered process variability than MOS transistors.

Design Goals

As a consequence of the population aging, chronic diseases are becoming the significant cause of death in most countries. Unfortunately, intermittent and asymptomatic features of many chronic disorders turn the task of diagnosis into a real challenge. Due to the new trends of wearable devices for biosensing, predicting and detecting the exact moment that chronic disease is happening became a possible solution. Hence, it can do a better and timely diagnostic. Therefore, in this project, we propose an ultra-low-power analog front-end to integrate a system for long-term monitoring. We can apply this work to several bio-signals, such as ECG, EMG, and EEG.

Block Diagram

Schematics

Low Power Low Noise Amplifier [1]
Variable Gain Amplifier (VGA)

 

High-Pass/Low-Pass Filter

 

Target Performance Summary

VDD 1.8 V
Power 5 uW
Input Noise 10 uV RMS (@ 0.05-100 Hz)
Offset Voltage 1 mV
Gain 10-100
Bandwidth 0.5-100 Hz
THD 0.1 % @ 1 Vpp

 

References

[1] Harrison, Reid R., et al. "A low-power integrated circuit for a wireless 100-electrode neural recording system." IEEE Journal of Solid-State Circuits 42.1 (2006): 123-133.

[2] Hogervorst, Ron, et al. "A compact power-efficient 3 V CMOS rail-to-rail input/output operational amplifier for VLSI cell libraries." IEEE journal of solid-state circuits 29.12 (1994): 1505-1513.

[3] Arnaud, Alfredo, Rafaella Fiorelli, and Carlos Galup-Montoro. "Nanowatt, sub-nS OTAs, with sub-10-mV input offset, using series-parallel current mirrors." IEEE Journal of Solid-State Circuits 41.9 (2006): 2009-2018.

[4] Krummenacher, Francois, and Norbert Joehl. "A 4-MHz CMOS continuous-time filter with on-chip automatic tuning." IEEE Journal of Solid-State Circuits 23.3 (1988): 750-758.

[5] Serra-Graells, Francisco, and Jose Luis Huertas. "Sub-1-V CMOS proportional-to-absolute temperature references." IEEE Journal of Solid-State Circuits 38.1 (2003): 84-88.

Team Members

B.Eng. Deni Germano Alves Neto

M.Sc. João Vitor Testi Ferreira

M.Sc. Luís Henrique Rodovalho

M.Sc. Rafael Sanchotene Silva

M.Sc. Thiago Daros Fernandes

Organization URL

https://ufsc.br/

Description

This design contains an analog signal process block that filters DC inputs, amplifies the AC components with an user selected voltage gain, and filters frequencies outside the bio-signals bandwidth, while consuming only 5 uW for a 1.8 V supply voltage.